Abstract
In a PET study, shortening of transmission scan time is highly desired for improving patient comfort and increasing scanner throughput. It necessitates a method that reduces statistical noise in attenuation correction factors (ACFs). The authors have evaluated non-linear Gaussian (NLG) filtering for smoothing transmission images reconstructed with filtered back-projection instead of using iterative reconstruction and segmentation methods. The NLG filtering operation is a variation of local weighted averaging in a neighborhood around a pixel, which weights are determined according to both distance in location and difference in pixel value. Several filtering steps with different NLG parameters can effectively reduce noise without losing structural information. The NLG smoothed transmission images are then forward projected to generate ACFs. Results with phantom and patient data suggested that the NLG filtering method is useful for attenuation correction using count-limited transmission data for both brain and whole-body PET studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.