Abstract

Small diameter implantable cardioverter-defibrillator (ICD) leads are prone to cause unexpected failures in remote phase. We demonstrate a case of unusual insulation breaks detected 5 years after ICD implantation. A 37-year-old male of Brugada syndrome received a prophylactic ICD implantation of a single-chamber device (AtlasVR, SJM). A dual coil ICD lead (Riata 1570, SJM) was implanted at right ventricular apex with passive fixation. Four years after, we found some non-reproducible epidodes of oversensing only when the patient press his hands together. We set the detection interval longer, and reduced the R wave sensitivity in order to prevent inappropriate shocks. After one year, we recognized intermittent episodes of noise oversensing during the interrogation. The ICD lead profile appeared to be normal though the lead impedence had gradually reduced (560 ohm to 350 ohm). Interestingly, the chest X-ray revealed the lead was torn in the longitudinal direction and some cables were seperated from the main body. He underwent a new lead implantation and has been fine without any complications. Since the lead profile remains within normal limits, and the lesion of insulation break is peculiar, SJM-Riata lead failure would be prone to be overlooked.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.