Abstract

Identification of homocystinuric newborns is hindered by the pitfalls of neonatal screening programs. We propose a fluorimetric HPLC method with a rapid pre-analytical step for homocysteine determination from neonatal dried blood spot cards. Homocysteine in blood spots sampled among 2000 healthy newborns on living day 4, averaged 2.92±2.07 μ M (range 0.4–7.5). In eight homocystinuric control children, mean values were 61.71±52.84 μ M (range 18.9–145.7). The method showed a good linearity ( r=0.999), precision (RSD<7%) and recovery (95%). The correlation between blood spots and plasma samples was r=0.90. This method has all the essential features for a homocystinuria screening program: an easy and rapid pre-analytical step combined with method linearity and precision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.