Abstract

To determine the normal SpO2 in healthy term newborns at mild altitude (MA, 780 metres) compared with sea level (SL), within the context of universal screening for critical congenital heart disease (CCHD). We studied 199 (119 at MA and 80 at SL) consecutively born healthy newborns. SpO2 recordings were at 24-72 h using Masimo SET Radical-7 on the right hand and left foot. Mean SpO2 was lower at MA compared with SL in the right hand (97.86 ± 1.58 vs 98.28 ± 1.41, p = 0.05) and left foot (98.49 ± 1.35 vs 98.90 ± 1.16, p = 0.03). No infant with SpO2 <95% had CCHD. Extrapolating with predicted regression lines set at 95% CI, a SpO2 cut-off of 95% would result in up to 3.5 times more false-positive screens at MA compared with SL. At MA, SpO2 is approximately 0.4% lower compared with SL. Our study supports the AAP recommendation suggesting algorithm cut-offs may need adjustment in high-altitude nurseries and suggest broadening it to MA as well.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.