Abstract

Conventional mano-volumetric techniques cannot measure changes in circumferential dimensions at several axial positions within a bowel segment. Our aims were to validate a new impedance planimetry probe for simultaneously measuring the cross-sectional area (CSA) at five axial positions in vitro and in vivo in 10 anesthetized pigs. The day-to-day coefficient of variation (CV) for CSA measured by the probe in cylindrical tubes of known diameter was 0.8-9.5%. The mean from actual diameter deviation ranged from 2.3 to 6.7%. In a conical tube the day-to-day CV was 2.3-8% and mean percentage deviation -2.8 to 1.0. Interposed narrowing sections caused a total CV of 7-13%. In vivo studies revealed variations in CSA, associated with expulsion of flatus. It is concluded that impedance planimetry allows simultaneous measurement of CSA at several levels within the rectum. In vitro validity was acceptable and alterations in lumen diameter were identified in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.