Abstract

Stent dislodgment from the delivery catheter is a well-known complication of angioplasty with stent implantation. The aim of our study was to investigate the feasibility, effectiveness, and safety of a new technique of intracoronary stent implantation in order to avoid stent loss in the intravascular space. Fifty consecutive patients were candidates for angioplasty and stent placement. During angioplasty, a technique was followed according to which the guide wire “hindered” the dislodgment of the stent from the balloon catheter. Successful angioplasty and stent placement were performed in 46 out of 50 patients (92% success rate). In three cases of failure of stent implantation, the stent moved onto the balloon catheter; however, this was impeded by the guide wire. One non-Q-wave myocardial infarction occurred. No major complications (Q-wave myocardial infarction, CABG, or death) were observed. In conclusion, the technique applied was feasible enough, safe, and effective. However, the appropriate modification of its “hardware” will render it even more feasible and user-friendly. Cathet. Cardiovasc. Diagn. 45:183–187, 1998. © 1998 Wiley-Liss, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.