Abstract

The neutron sensitivity of vanadium self powered neutron detectors (Type 5503-B), manufactured by Studsvik, Sweden, has been measured. Using the calculational model developed by Warren, the neutron sensitivity of these detectors has been computed taking into account additional factors viz. flux depression caused by the detector and the interaction of 52V gamma rays with the emitter, the correction factors due to which have been evaluated to be 0.957 and 1.03 respectively. The evaluation of different parameters that are necessary in this connection and the neutron flux density determination at the position of irradiation of these detectors by gold foil activation, are presented. The measured and calculated values of neutron sensitivity are observed to agree to within 4%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.