Abstract
The neutron dose equivalent rate (DR) leaking from the self-shielded cyclotron was measured using an FHT-751 neutron counting system in the facility at the Cyclotron Center of Chung Shan Medical University. This system was calibrated using two (252)Cf neutron sources and simulated according to MCNP code. The results show various DRs up to 120 microSv h(-1) in the cyclotron room. Two-dimensional distributions of measured neutron DRs indicate an explicit, heavy leakage of neutrons through the self-shielded interlock. The neutron DR of the operating cyclotron that is hazardous to the health of medical personal and the public is evaluated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.