Abstract

The aim of this experimental study was to investigate the effect of Nd:YAG laser on metallic tracheobronchial stents, because Nd:YAG laser could be used for resection of tumourous tissue growing through stent meshes (Strecker-, Palmaz-, Wallstent). An in vitro worst-case-model was used to investigate the effects of different types of laser application on metallic stents: non-contact application, and contact application with bare fibres and with hemispherical sapphires. In the non-contact method and with bare fibres stent destruction occurred at power settings exceeding 10 W, whereas in the contact method with hemispherical sapphires power settings of 20 W were possible without stent damage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.