Abstract
Radiographic confirmation of enteral feeding tube placement is a common practice representing considerable expense and causing delay in the initiation of enteral nutrition therapy. We evaluated an enteral feeding tube with a pH sensor, which allows immediate verification of the location of the tube by assessment of the pH upon insertion. Insertion pHs were obtained for 24 intensive care unit patients requiring feeding tube placement. Placement was verified radiographically and compared with expected location on the basis of the pH profile. The radiograph and the insertion pH profile were in agreement in 87.5% (21 of 24) of the cases. Concomitant use of histamine blockers did not affect the ability of the pH sensor to detect placement accurately (Fisher's Exact Test, p 5.71) Use of these pH measurements eliminates the need for radiographic documentation of placement, provides a savings for the patient, and may be beneficial in promoting enteral feedings in critically ill patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.