Abstract

The magnetic particle imaging (MPI) method directly images the magnetization of super-paramagnetic iron oxide (SPIO) nanoparticles, which are contrast agents commonly used in magnetic resonance imaging (MRI). MPI, as originally envisioned, requires a high-bandwidth receiver coil and preamplifier, which are difficult to optimally noise match. This paper introduces Narrowband MPI, which dramatically reduces bandwidth requirements and increases the signal-to-noise ratio for a fixed specific absorption rate. We employ a two-tone excitation (called intermodulation) that can be tailored for a high-Q, narrowband receiver coil. We then demonstrate a new MPI instrument capable of full 3-D tomographic imaging of SPIO particles by imaging acrylic and tissue phantoms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.