Abstract

A small ionization needle with an ultrasharp, ultrafine tip is introduced. It is lab-fabricated from tungsten wire and serves as a corona discharge emitter in nanoelectrode atmospheric pressure chemical ionization mass spectrometry (nAPCI-MS). Tip radii ranged from 8 to 44 nm, up to 44× smaller than the sharpest previously reported corona needle. Because of this, nAPCI was able to operate at +1.0 kV with no auxiliary counter electrode. Alternatively, at +1.2 kV, nAPCI could be enclosed in a small plastic assembly for headspace analysis with a sampling tube attachment as long as 15 m. No added heat or gas flow was necessary. The efficacy of nAPCI-MS was demonstrated through needle durability studies and direct analysis of vapors from real-world samples. Provisional identifications include ibuprofen from a pharmaceutical tablet, albuterol aerosol sprayed from a medical inhaler, cocaine from paper currency, caffeine from a fingertip, and bisphenol E from a paper receipt.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.