Abstract
A wearable glucose sensor for tear sugar measurement was fabricated and evaluated. The sensor was constructed by immobilizing glucose oxidase onto a flexible oxygen electrode, which was fabricated using "Soft-MEMS" techniques onto a functional polymer membrane. Linear relationship between glucose concentration and output current was obtained in a range of 0.049 to 1.37 mmol/l, with a correlation coefficient of 0.998. In addition to characteristics evaluation, the sensor was utilized to tear glucose measurement of a Japan white rabbit. The change of tear sugar induced by oral-administration of glucose was monitored as a current change of the wearable glucose sensor attached on the eye site. In this investigation, the tear glucose level varied from 0.3 to 0.93 mmol/l.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.