Abstract

The Compton detector, with fast time response, linear range over many orders of magnitude and low sensitivity to X-ray fluxes, can be used to measure the dose of time resolved X-ray, whose sensitivity is decided by the materials and dimensions of its collector, dielectric and metal case. Since the detector dimension is energy dependent, there will be an optimal dimension for a detector to measure bremsstrahlung X-ray dose. Monte-Carlo method was used to design a Compton detector which will be used to measure the bremsstrahlung X-ray dose produced by the Dragon-Ⅰ accelerator. The bremsstrahlung spectrum of Dragon-Ⅰ was computed by MCNP program, as well as the Compton electron number as a function of plexiglass thickness, aluminum case thickness and plumbum collector thickness. According t

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.