Abstract

This article presents a novel design for magnetic resonance imaging (MRI) gradient systems. This design may allow the development of MRI scanners that are capable of imaging large regions with high performance while minimizing the potential for nerve stimulation. The general concept of the gradient system is that spatial oscillation is incorporated such that each gradient coil creates multiple, approximately linear gradient regions that oscillate in gradient polarity. Separate radiofrequency (RF) coil arrays are designed to be sensitive to the signals within each linear region and thus allow signal measurements to be obtained separately from each region. Enabling image acquisition in the transition region that separates each pair of adjacent linear regions requires a second gradient system with imaging regions that overlap and coincide with the transition regions of the first gradient system. Imaging the extended field of view (FOV) is accomplished by interleaved operation of the two gradient systems. Simulated annealing is used to create designs for both longitudinal and transverse gradient systems with two imaging regions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.