Abstract
A simple formula for the spatial resolution of transmission proton radiography is derived for two different methods of measuring the proton coordinates. The effect of multiple Coulomb scattering and energy loss are taken into account. Experimental measurements of the spatial resolution have been done and are compared with the calculations. The technique of measuring entrance and exit coordinates in coincidence for each single proton improves the spatial resolution by a factor of 8 compared to a single coordinate measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.