Abstract

The National Institute of Standards and Technology (NIST), Advanced Surface Microscopy (ASM), and the National Metrology Centre (NMC) of the Agency for Science, Technology, and Research (A*STAR) in Singapore have completed a three-way interlaboratory comparison of traceable pitch measurements using atomic force microscopy (AFM). The specimen being used for this comparison is provided by ASM and consists of SiO2 lines having a 70-nm pitch patterned on a silicon substrate. For this comparison, NIST used its calibrated atomic force microscope (C-AFM), an AFM with incorporated displacement interferometry, to participate in this comparison. ASM used a commercially available AFM with an open-loop scanner, calibrated with a 144-nm pitch transfer standard. NMC/A*STAR used a large scanning range metrological atomic force microscope with He-Ne laser displacement interferometry incorporated. The three participants have independently established traceability to the SI (International System of Units) meter. The results obtained by the three organizations are in agreement within their expanded uncertainties and at the level of a few parts in 104.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.