Abstract

The use of high-resolution Ge(Li) detectors in non-destructive instrumental neutron activation analysis (INAA) of mineral materials makes possible the simultaneous determination of a number of trace elements. In routine applications of INAA the use of a multi-element standard (MES) has proved to have advantages over a set of standards for each determined element. An MES has been prepared containing 8 trace elements mixed in a suitable proportion and giving, after neutron activation, long-lived γ-emitters, the γ-ray lines of which regularly occur in the γ-ray spectra of silicate mineral materials. This method has been used in the determination of Sc, Cr, Co, Rb. Cs, Eu, Hf and Th in samples of standard rocks and moldavites.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.