Abstract

In-vitro and In-vivo evaluation of feasibility and safety of MRI of the brain at 1.5 T in patients with implanted pacemakers (PM). 24 PM models and 45 PM electrodes were tested In-vitro with respect to translational forces, heating of PM leads, behaviour of reed switch (activated vs. deactivated) and function at a 1.5 T MRI-system (actively shielded, maximum field gradient: 30 mT/m; rise time: 150 T/m/s). Based on these results, 63 MRI examinations in 45 patients with implanted PM were performed. Prior to MRI the PM were re-programmed in an asynchronous mode. The maximum SAR of MRI-sequences was limited to 1.2 W/kg. Continuous monitoring of ECG and pulse oximetry was performed during MRI. PM inquiry was performed prior to MRI, immediately after MRI and -- to assess long-term damages -- three months after the MRI exams, including determination of stimulation thresholds to assess potential thermal myocardial injuries at the lead tips. Translational forces (F (max) < or = 560 mN) and temperature increase (DeltaT (max) < or = 2.98 degrees C) were in a range which does not represent a safety concern from a biophysical point of view. No changes to the programmed parameters of the PM or damage of PM components were observed neither In-vitro (n = 0/24) nor In-vivo (n = 0/63). Despite the strong magnetic field, the reed switch remained deactivated in 54 % (13/24) of the cases during In-vitro simulated MRI exams of the brain. All patient studies (n = 63/63) could be completed without any complications. Atrial and ventricular stimulation thresholds (expressed as pulse duration at 2-fold rheobase) did not change significantly immediately post-MRI nor in the 3 months follow-up (pre-MRI: 0.17 ms +/- 0.13 ms, post-MRI: 0.18 ms +/- 0.14 ms, 3 months follow-up: 0.17 ms +/- 0.12 ms). MRI of the brain at 1.5 Tesla can be safely performed in carefully selected clinical circumstances when appropriate strategies are used (re-programming the PM to an asynchronous mode, continuous monitoring of ECG and pulse oximetry, limiting the SAR value of the MRI sequences, cardiological stand-by). Based on these studies, implanted PM should not longer be regarded as an absolute contraindication for MRI at 1.5 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.