Abstract

Our aim was to evaluate whether it is possible to visualize slow flow within a small catheter placed inside a living animal. We used a flow-sensitive, single-shot turbo spin-echo (SS-TSE) MRI sequence, developed in house, based on diffusion-weighted (DW) techniques. Four anesthetized pigs were used as models. A plastic catheter was surgically placed within the common bile duct (CBD). To mimic flow, the catheter was filled with Ringer's acetate and connected to a pump. b factors (s/m(2)) of 0, 6, and 12, with flow velocities raging from 0 to 1.32 cm/s, were used. A total of 375 images were obtained and examined. After correction for bowel movement artifacts, all images displayed the catheter on zero flow. With a flow of 0.66 cm/s or higher, no images displayed the catheter with a b factor of 6 or 12. On the slower flow velocities, it was variable whether the catheter was visible or not, but at b=6 and flow 0.17 cm/s all catheters were viewable. This method made it possible to perform a semiquantitative evaluation of flow velocities in vivo, dividing flow into three groups.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.