Abstract

Magnetic resonance imaging (MRI) has unique advantages for guiding interventions, but the narrow space is a major challenge. This study evaluates the feasibility of a remote-controlled hydrostatic actuator system for MRI-guided targeted needle placement. The effects of the hydrostatic actuator system on MR image quality were evaluated. Using a reference step-and-shoot method (SS) and the proposed actuator-assisted method (AA), two operators performed MRI-guided needle placement in targets (n = 12) in a motion phantom. The hydrostatic actuator system exhibited negligible impact on MR image quality. In dynamic targets, AA was significantly more accurate and precise than SS, with mean ± SD needle-to-target error of 1.8 ± 1.0 mm (operator 1) and 1.3 ± 0.5 mm (operator 2). AA reduced the insertion time by 50% to 80% and total procedure time by 25%, compared to SS. The proposed hydrostatic actuator system may improve accuracy and reduce procedure time for MRI-guided targeted needle placement during motion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.