Abstract
To evaluate improvements in image homogeneity in pelvic MR imaging at 3 Tesla (T) using two different dielectric pads. A total of eight healthy females were scanned using a 3T MR scanner equipped with a body-array coil. Axial and sagittal fast spin-echo T2-weighted images (T2WI) (TR/TE = 3200 msec/94 msec), axial fast spin-echo T1-weighted images (T1WI) (TR/TE = 700 msec/11 msec), and sagittal half-Fourier acquisition single-shot turbo spin-echo (HASTE) images (TR/TE = 3000 msec/100 msec) were performed for pelvic imaging. Sequences were repeated with dielectric pads (consisting of either ultrasound [US] gel or water), and without pads. Three or four regions of interest (ROIs) were placed on fatty tissues and the ratio of minimum to maximum signal intensity (RSI) was calculated as a marker of image homogeneity. RSI was significantly higher on T2WI and T1WI when using dielectric pads than when no pad was used. A similar tendency was observed in RSI on HASTE. No significant difference was found between images with US gel pads and those with water pads. Dielectric pads consisting of either US gel or water are effective in improving image homogeneity of the pelvis on 3T MRI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.