Abstract

The authors analyzed the accuracy and precision of focal-spot placement with a magnetic resonance-guided, focused-ultrasound system. Average absolute accuracy errors ranged from 0.2 to 1.0 mm, and average absolute individual precision errors ranged from 0.2 to 0.3 mm. To prevent damage to vital structures, single sonications and sonication grids should be placed beyond approximately 2 and 1 mm, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.