Abstract
It is often required in neutron calibration fields to determine the dose of gamma-rays produced in a neutron source and in the surroundings, because some neutron detectors are sensitive both to neutrons and gamma-rays. As a first step estimating neutron and gamma-ray energy spectra and fluences simultaneously in a neutron field, the gamma-ray response function for a 3 He proportional counter was calculated using EGS4/PRESTA code. In the simulation, the gas multiplication factor, which depends on the detection position in the gas region along an anode wire, was taken into account to obtain the precise response function. For comparison, we measured response functions using different reference gamma-ray sources such as 241 Am , 137 Cs , and 60 Co , which were located vertically 55 mm away from the center of the anode wire. Calculated response functions agreed well with experimental ones. Monte Carlo code EGS4 is thus useful for the simulation of gamma response functions for neutron counters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.