A simple modification to the spark shield of the ion source of a spark source mass spectrograph is discussed. This helps to maintain a low partial pressure of residual gases during sparking of reactive materials. A grid is placed over the viewing slit to prevent energetic neutrals and ions from striking the walls of the source chamber, thereby desorbing unwanted gases into the ion source chamber. The results show a reduction of an order of magnitude in the total pressure during sparking over the conventional spark shield.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.