Abstract

ABSTRACTA butanol-type ultrafine condensation particle counter (UCPC, Model 3776, TSI, Inc., Shoreview, MN, USA), which can achieve a 50% detection efficiency diameter (d50) of 2.5 nm using a capillary-sheath structure, was modified and tested in the laboratory for airborne measurements. The aerosol flow rate through the capillary is a key factor affecting the quantification of aerosol particle number concentrations. A pressure-dependent correction factor for the aerosol flow rate was determined using a newly added mass flow meter for the sheath flow and the external calibration system. The effect of particle coincidence in the optical sensing volume was evaluated using an aerosol electrometer (AE, Model 3068B, TSI, Inc.) as a reference. An additional correction factor for the coincidence effect was derived to improve the quantification accuracy at higher concentrations. The particle detection efficiency relative to the AE was measured for mobility diameters of 3.1–50 nm and inlet absolute pressures of 10...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.