Abstract

The purpose of this study was to analyze the dynamic behavior of a dental implant with a stress-absorbing element, using dynamic analysis. Two model types, stress-absorbing model with a resilient stress absorber made of polyoxymethylene and non-stress-absorbing model with rigid titanium, were employed. In both model types, the implant was 4.0 mm in diameter and 13.0 mm in length and placed in the mandibular first molar region. Shapes of the finite element implant and implant-bone were modeled using computer-aided design. All calculations for the dynamic analysis were performed using the finite element method. It was found that the stress-absorbing model had a lower natural frequency than the non-stress-absorbing model. In addition, the stress-absorbing model had a higher damping effect than the non-stress-absorbing model. It was concluded that mode superposition transient dynamic analysis is a useful technique for determining dynamic behavior around dental implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.