Abstract

We report the case of a severe prosthetic aortic stenosis in a 61-year-old male patient with an aortic Mitroflow LF bioprosthesis (Sorin Group) at early (6 months) postoperative echocardiographic follow-up. At reintervention, we found significant stent deformation, asymmetric orientation of the posts, and subsequent central kinking and prolapse of one leaflet. Maldistribution of pledgeted mattress sutures over the flexible stent was found to be the origin of its permanent deformation. Simple technical preventive strategies of this previously unreported complication are suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.