Abstract
The purpose of the study was to determine the optimal number of throws to ensure knot security. Knots were tied with 3, 4, 5, or 6 square throws with 0-gauge coated polyester, polydioxanone, polypropylene, and polyglactin 910. The suture was soaked in 0.9% sodium chloride and subsequently transferred to a tensiometer and broken. A total of 225 knots were tied. Regardless of the suture type, tension at failure for knots with 4 throws, 5 throws, and 6 throws was higher than tension at failure of knots with only 3 throws (p < 0.05 for each). We found no difference in the tensile strength between knots with 4, 5, or 6 throws (p > 0.05 for each). Knots with 4 throws were significantly more likely to come untied than knots with 5 or 6 throws (p < 0.01). Under laboratory conditions, the ideal knot has 5 throws to maximize tensile strength and rate of untying. This finding does not seem to vary by type of suture material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.