Abstract

Because the Nellcor N-100 pulse oximeter is generally quite sensitive to electrocautery interference, erroneous oxygen saturation values may be displayed and an audible false alarm may sound. We have discovered, through trial and error, several techniques that may minimize both the interference and the false alarms. Keeping the pulse oximeter sensor and the oximeter unit as far as possible from the surgical site and the electrocautery grounding plate will usually decrease the interference. If this step is inadequate, the incidence of false alarms can be decreased by setting the high pulse rate alarm to approximately 160, operating the unit in mode 2 (the rapid-response mode), and setting the "alarm off" function to 120 seconds. These steps will usually markedly decrease the incidence of false alarms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.