Abstract

A new method of reducing the amount of reagent and sample for determination of thrombomodulin (TM) was developed based on competitive immunoreaction using a portable glucometer (PGM). Two types of nanocomposites, TM protein-modified magnetic nanoparticles (MNPs-TM) and TM antibody-/glucose oxidase-modified gold nanoparticles (Ab-GNPs-GOx), were prepared. Their binding product, MNPs-TM-Ab-GNPs-GOx, in the microvolumetric solution was used to catalyze the oxidation of glucose, leading toa decline of the glucose content. The TM-involved competitive immunoreaction had a negative effect on the generation of MNPs-/GNPs-based nanocomposites and inhibited the catalytic oxidation of glucose. The glucose content difference in the microvolumetric solution, which was revealed by a PGM, was in proportion to the logarithm of the TM concentration from 25ngmL-1 to 2.5μgmL-1. The limit of detection was 5.7ngmL-1. Microvolumetric solution and a PGM were used in the measurement, which overcame some deficienciesof classical methods in chemo/biosensing, for example, special instrument, complicated measurement procedure, and high cost.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.