Abstract

Four position sensitive proportional counters will be delivered by the Danish Space Research Institute as focal plane instruments for the Soviet Danish Röntgen Telescope (SODART). These detectors will incorporate the novel microstrip electrode design. This design has been tested with xenon gas and an 55Fe X-ray source. The energy resolution is better than 14% FWHM up to gas gains of 10 4, the rise time of pulses from X-ray events is fast, 100 ns, allowing for efficient background rejection by rise time analysis. A position resolution of 1 mm FWHM is easily obtained by resistive charge division of the signals from the cathodes and from a wire grid placed in front of the microstrip plate. The detector performs without significant degradation of gain or energy resolution at fluxes exceeding 10 6/s cm 2 and a total dose of 10 11 photons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.