Abstract

Pulsed laser systems produce spectroscopic signals with large shot-to-shot variations, short durations, and low duty cycles. The most satisfactory signal processing system is gated integration, with pulse averaging and ratioing to a reference signal. We describe a system based on commercially available modular integrators and a hobbyist microcomputer which performs the averaging and ratioing digitally. System performance is equal to or better than that of fully analog systems, while the cost is low and construction is straightforward.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.