Abstract

The aim of this study was to evaluate the performance of chemiluminescence immunoassays for anti-GAD65 and anti-insulin antibodies following user verification guidelines. The analytical performance of anti-GAD65 and anti-insulin antibodies using a MAGLUMI 2000 analyzer was verified following user verification guidelines by the Clinical and Laboratory Standards Institute. Performance specifications including precision, linearity, carry-over, cutoffs for positive results, reference intervals, and comparability with pre-existing commercially available radioimmunoassays using patient specimens and certified reference material were verified (coefficients of variation for precision of anti-GAD65 and anti-insulin antibodies were 2.6% and 3.4%, respectively). Comparability assessed using clinical serum specimens showed overall agreement with radioimmunoassay of 87.2% (95% confidence interval 74.8% - 94.0%) for the anti-GAD65 antibody assay and 85.4% (95% confidence interval 71.6% - 93.1%) for the anti-insulin antibody assay. The results of this study verified the analytical performance of MAGLUMI anti-GAD65 and anti-insulin antibody assays for clinical use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.