Abstract

This study proposes a method for precise measurement of time waveforms of minute sinusoidal displacements using continuous waves. The technique involves measuring the S-parameter of ultrasonic waves reflected from an object undergoing sinusoidal vibration with a network analyzer. When the sinusoidal displacement is not sufficiently small, second-order side peaks emerge in its inverse Fourier transform, alongside the main and first-order side peaks. This method, which incorporates second-order side peaks previously overlooked, enhances the accuracy of measurements. It facilitates precise measurement of sinusoidal displacements ranging from 0.3 to 50 μm at frequencies between 10 and 300 Hz.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.