Abstract

A simple and precise RP‐HPLC method was developed and validated for the determination of pioglitazone hydrochloride in pharmaceutical dosage forms. Chromatography was carried out using Kromosil- C18 ODS column (250 x 4.6 mm; 5 μm), mixture of acetate buffer: methanol (40:60 v/v) as the mobile phase at a flow rate 1.0 mL/min. The analyte was monitored using UV detector at 254 nm. The retention time for pioglitazone HCl was 3.063 min. The proposed method was found linear in the concentration range of 20.0‐70.0 μg/ml with correlation coefficient of r2=0.9999. The developed method has been statistically validated and found simple and accurate. The mean recoveries obtained for pioglitazone HCl were in the range 99.20-101.59%. Due to its simplicity, rapidness, high precision and accuracy of the proposed method it may be used for determining pioglitazone HCl in bulk and dosage forms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.