Abstract

Increased wear leads to elevated systemic and local metal ion concentrations for patients treated with metal-on-metal bearings. The local metal ion content in the close environment of the joint replacement (e.g. joint aspirate or tissue) is several times higher compared to the systemic metal content (e.g. in blood or serum). As a result of increased metal ion levels, local and systemic effects, such as osteolysis, pseudotumors, sensitization or in rare cases toxicity may occur. Although the definition of a specific threshold to define clinical problems is difficult due to a lack of sensitivity, the systemic metal concentration is frequently measured clinically. Currently a threshold for cobalt and chromium between 4µg/l and 7µg/l is under debate. Very high levels (≥ 20µg/l) or a steady increase over time should be a warning sign; however, metal ion levels should not be interpreted as a single diagnostic tool but rather in the entire context of the clinical, radiological and cross-sectional imaging, metal artefact reduction sequence (MARS) magnetic resonance imaging (MRI), ultrasound and computed tomography (CT) findings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.