Abstract

Preoperative and postoperative ion concentrations were measured in 29 metal-on-metal, large-diameter head total hip arthroplasty (LDH-THA) patients. Mean chromium, cobalt (Co), and titanium levels from LDH-THA were 1.3, 2.2, and 2.7 μg/L at 12 months. The open femoral head design showed significantly higher Co concentrations than the closed design (3.0 vs 1.8 μg/L, P = .037). Compared with previously published ion levels from a hip resurfacing system presenting the same bearing characteristics, Co levels were significantly higher in LDH-THA (2.2 vs 0.7 μg/L, P < .001). This study has demonstrated that the addition of a sleeve with modular junctions and an open femoral head design of LDH-THA causes more Co release than bearing surface wear (157% and 67%, respectively). Even if no pathologic metal ion threshold level has been determined, efforts should be made to minimize its release. We recommend modification or abandonment of the modular junction and femoral head open design for this specific LDH-THA system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.