Abstract

This article determines the accuracy and reliability of dual-energy computed tomography (DECT) with metal artifact reduction (MAR) in the evaluation of femoral component rotation after total knee arthroplasty (TKA), in comparison with conventional CT images. A total of 49 patients (mean age, 69 years; 42 women) who underwent TKA between January 2019 and March 2020 were retrospectively enrolled. Femoral component rotation, including the anatomic and surgical transepicondylar axes, was evaluated with preoperative conventional CT and postoperative conventional CT and DECT with MAR. Surgical femoral component rotation was also assessed as a reference standard. Accuracy was assessed using paired t-test, and inter- and intraobserver reliability using intraclass correlation coefficients (ICCs) based on postoperative conventional CT and DECT with MAR. Clinical outcomes were evaluated using the Knee Society objective and functional scores. Accuracy of femoral component rotation was not significantly different from that of surgical rotation with both conventional CT and DECT with MAR. However, inter- and intraobserver reliability were better for DECT with MAR (ICC: 0.953-0.966) than for conventional CT (ICC: 0.641-0.749). The Knee Society objective and functional scores improved 1 year postoperatively. CONCLUSION: DECT with MAR showed accurate and more reliable results than did conventional CT in the evaluation of femoral component rotation after TKA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.