Abstract

One-compartment bulk electrolysis and simultaneous spectroscopic measurements are realized in a conventional spectroscopic cuvette without separator by using a mesh-type working electrode with extremely large surface area and a wire-type counter electrode with very small surface area. Spectrophotometric monitoring revealed complete electrolysis in a first-order kinetics. This technique was applied to mediated titration of cytochrome c and bilirubin oxidase for determining their redox potentials. Kinetics for the solution redox reaction between protein and mediator is described. The subtraction of spectral background due to mediator adsorption is very easy because of high reproducibility. The experiments can be done under completely anaerobic conditions. Low-absorbance protein samples (of low concentrations or small absorption coefficients) and hydrophobic proteins (such as membrane-bound proteins) are acceptable for measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.