Abstract

Medial lag screw migration in Gamma nail system is an uncommon complication. In 92-year-old woman, the medial lag screw migration into the pelvic with nonunion oftrochanteric fractureat 4 months after the osteosynthesis. The computed tomography scan did not reveal intestinal organs injury by the protruded screw and the removal of the implant eased her right hip pain. During the operation, the lag screw migration by the incorrect engagement of the set screw outside of the groove on the lag screw was observed, which was considered the reason. To avoid such a hazardous complication in Gamma nail system for trochanteric fracture, correct engagement of the set screw into the groove on the lag screw should be confirmed after tightening the set screw.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.