Abstract

Objective: To evaluate the Ulson intramedullary fixation method, with external fixation in which the level of the external locking of the Kirschner wires is varied, and without external fixation. Methods: Eighteen porcine tibias were used. Transverse osteotomy was performed in the region of the tuberosity, and two intramedullary Kirschner wires were inserted into each specimen, using three different assembly patterns: Group I: locking with external minifixator at a height of 3.0 cm; Group II: locking at a height of 4.5 cm; Group III: without external locking. Mechanical shear tests were then conducted, to determine the maximum load, proportionality limit and coefficient of rigidity. Results: There were no significant differences in maximum load or proportionality limit between the groups. The coefficient of rigidity was higher in Group II. Conclusion: The locking height for the Kirschner wires in Ulson's method, within the limits evaluated, did not harm the stability of the fracture fixation system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.