Abstract

The aim of this study was to characterize the mechanical and self-expansion properties of braided bioabsorbable stents. In total four different stents were manufactured from PLLA fibres using a braiding technique. The changes in radial pressure stiffness and diameter recovery of the stents were determined initially, and after insertion and release from a delivery device. The braided stents were compared to three commercially available metallic braided stents. The changes in physical and mechanical properties of the PLLA fibres and stents during in vitro degradation were investigated. After release from the delivery device, the PLLA stents did not fully recover to their original diameter. The radial pressure stiffness of the bioabsorbable stents was similar to that of the metallic stents. The in vitro degradation study showed that the stents would keep at least half of their initial radial pressure stiffness for more than 22 weeks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.