Abstract

The concentration of oxygen delivered by a Bird Mark 7 respirator with the air mixture control adjusted to full air dilution was evaluated under clinical circumstances. A mixture of carbon dioxide in air was substituted for oxygen to drive the respirator because the concentration of carbon dioxide in the inspired air could be measured by an analyzer with a short response time. One hundred and three single inspirations of 7 participants were analyzed for inspired concentration of carbon dioxide tracer gas; for peak and mean flows; and for pressure, volume, and cycling time. The concentration of delivered gas was found to be directly proportional to the pressure (P) and inversely related to the mean flow () with the following relationship: inspired concentration = 58.2757 − 0.4567 + 0. 520 P. The concentration of inspired driving gas varied from 31 per cent to 76 per cent. This is less than the inspired concentration predicted from lung—thorax analogs, probably indicating that performance of respirators nee...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.