Abstract
To validate a fast gradient echo sequence for rapid (9 s) quantitative imaging of lung water. Eleven excised pig lungs were imaged with a fast GRE sequence in triplicate, in the sagittal plane at 2 levels of inflation pressure (5 and 15 cm H(2) O), an intervention that alters T(2) *, but not total lung water. Images were acquired alternating between two closely-spaced echoes and data were fit (voxel-by-voxel) to a single exponential to determine T(2) * and water content, and compared with gravimetric measurements of total water. T(2) * averaged 1.08 ± 0.02 ms at 5 cm H(2) O and 1.02 ± 0.02 ms at 15 cm H(2) O (P < 0.05). The measure was reliable (R(2) = 0.99), with an average mean error of 1.8%. There was a significant linear relationship between the two measures of water content: The regression equations for the relationship were y = 0.92x + 19 (R(2) = 0.94), and y = 1.04x + 4 (R(2) = 0.96), for 5 and 15 cm H(2) O inflation pressure respectively. Y-intercepts were not statistically different from zero (P = 0.86). The multi-echo GRE sequence is a reliable and valid technique to assess water content in the lung. This technique enables rapid assessment of lung water, which is advantageous for in vivo studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.