Abstract
Viscometers in the Faculty of Chemical Engineering, University Teknologi MARA, are subject to heavy utilization from the members of the faculty. Due to doubts surrounding their result integrity and maintenance management, Measurement System Analysis was executed. 5 samples of drilling muds with varied barite content from 5 – 25 weight% were prepared and their rheological properties determined in 3 trials by 3 operators using the viscometers. Gage Linearity and Bias Study were performed using Minitab software and the result shows high biases in the range of 19.2% to 38.7%, with non-linear trend along the span of measurements. Gage Repeatability & Reproducibility (Nested) analysis later produces Percent Repeatability & Reproducibility more than 7.7% and Percent Tolerance above 30%. Lastly, good and marginal Distinct Categories output are seen among the results. Despite acceptable performance of the measurement system in Distinct Categories, the poor results in accuracy, linearity, and Percent Repeatability & Reproducibility render the gage generally not capable. Improvement to the measurement system is imminent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IOP Conference Series: Materials Science and Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.