Abstract

Using an automated oven evaporation technique combined with the coulometric Karl Fischer method, the mass fraction of water has been measured in cement, coal, and refined oil samples. The accuracy of this method was established by using SRM 2890, water-saturated 1-octanol that was added to white oil. The samples were analyzed for total reactive Karl Fischer reagent (KFR) material, for interfering materials, and for material that does not react with the aldehyde-ketone KFR. All of the samples yielded volatile material that reacted with the standard KFR. None of the samples contained significant masses of material that reacted with iodine. The cement and coal SRMs contained no material that reacted with methanol and very little material that did not volatilize at 107 degrees C. The refined oils contained some material that was volatile at 107 degrees C and some at 160 degrees C. However, none of this material reacted with the aldehyde-ketone reagent. These results show that the material in the solid samples is water and that the material in the refined oils is a material other than water which reacts with methanol to form water.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.