Abstract
This paper presents the development of a Tunable Light Pulse Generator built to measure the quantum efficiency of photodetectors under continuous or pulsed (20ns FWHM) illumination as a function of wavelength in the Visible–UV region. For this purpose, a miniature Hamamatsu R5600U-04 PMT with low cathode resistivity can be used as reference detector. Dedicated electronics was developed in order to measure simultaneously the cathode and anode currents. This should allow to determine precisely the gain as a function of the high voltage applied to the PMT. Preliminary measurements using the TLPG do not demonstrate significant differences between the dynamic (∼10000 photons) and DC responses of PIN photodiodes down to 300nm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.