Abstract

A method for measurement of peak and total efficiencies for point sources measured on low-energy gamma-ray detectors is presented. The efficiencies are measured with radionuclides emitting gamma-rays in a two-step cascade or in a crossover transition. One of the photons emitted in the cascade must have an energy which is much lower than the other in order to attain stronger attenuation as compared to the higher energy photon. The method is illustrated by measurements of the efficiencies with 179Ta and 57Co.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.