Abstract

In the measurement of neutrons from the medical electron accelerator by a rem counter, two problems disturb accurate measurements. One is the pile-up of signals produced by X-rays during each X-ray burst and the other is the increased counting loss caused by bunched nature of yielded neutrons. The time spectrum of neutrons measured by the rem counter 2202 D (manufactured by Studsvik) rises up to a maximum value by about 20 microseconds and then falls down exponentially with a time constant of about 90 microseconds. On the other hand, that of X-rays is roughly rectangular with several microseconds width. A time discriminating system was prepared to be combined with the rem counter, which was triggered by leading edge of electron beam pulses, rejected pile-up signals due to X-ray bursts, and counted pulses of neutrons in a specified time window. The system discriminated the pile-up enough to measure neutrons at a X-ray dose rate of at least 30 mGy/h. Nonparalyzable counting loss correction was practicable upto about 10 mSv/h for the beam pulse rate of 85 Hz, in which the dead time of the rem counter was estimated as 4 microseconds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.